You are ventilating a full term baby with meconium aspiration syndrome, he was fine on the machine, suddenly the monitor shows severe drop in O2 saturation. You checked the ventilator; it was working well with its connections. Of the following, the MOST likely cause is
28 views
- O2 source supply
- endotracheal tube blockade (Correct Answer)
- self extubation
- low inspiratory pressure value
Explanation
B
All the given options are true causes for sudden drop of 02. The possible causes
of desaturating baby on ventilator are; Oxygen supply, machine and it is connections,
settings, tube block, self extubation and the patient himself. In the given scenario,
the most likely cause is B because of meconium aspiration. The best action to such scenario
is to disconnect the baby from the machine and use Ambu bagging gently to check the
air entry and rise of O2 saturation. if no rise you need to change the tube. In case of
unequal air entry, pneumothorax is another possibility.