You are starting the first case on a Sunday morning in the emergency theatre. After induction of general anaesthesia, despite adequate mask ventilation using 6L/minute of oxygen flow, the oxygen saturation begins to fall. The oxygen analyser at the common gas outlet (fuel cell) and at the mask end of the breathing system (paramagnetic analyser) reads inspired oxygen concentration as 21%. Despite turning the oxygen cylinder on (pressure reads 90 bar), the oxygen saturation continues to fall. The single most important next step in the management is:

22 views
  • Immediate tracheal intubation.
  • Ventilate using a resuscitation bag and auxiliary oxygen source from the same anaesthetic machine
  • Change the pulse oximeter probe.
  • Disconnect the oxygen pipeline. (Correct Answer)

Explanation

Correct Answer: D. Disconnect the oxygen pipeline.

It is very unlikely that both oxygen analysers (fuel cell and paramagnetic
analyser) are faulty. Therefore, for some reason 21% oxygen (air) is being
delivered to the patient. The most likely reason is the pipelines have been
swapped over and the oxygen pipeline is delivering airSome anaesthetic machines have a pipeline preference by setting the
cylinder pressure regulators to 350kPa (50 psi), in which case even if the
cylinder is turned on gases will be delivered from the pipeline. This feature
is incorporated to prevent the premature emptying of the cylinder when the
reserve cylinder is accidentally left turned on. Hence, disconnecting the
pipeline allows the anaesthetic machine to deliver oxygen from the reserve
cylinder.
Immediate tracheal intubation is not required, as the patient is being
adequately ventilated using a bag and mask. If the pipeline is not
disconnected, the auxiliary oxygen source will also deliver 21% oxygen