You are planning to evaluate three fluid warmers in a laboratory setting. You need to accurately monitor the temperature of the fluid and would like to detect a very small difference of up to 0.001°C. Which of the following temperature measurement devices is most suitable for this research?

26 views
  • Mercury thermometer.
  • Infrared thermometer.
  • Platinum resistance thermometer. (Correct Answer)
  • Thermistor.

Explanation

Correct Answer: C. Platinum resistance thermometer.

A resistance thermometer displays a linear increase in resistance with
increasing temperature. Over the range of 0-100°C, the change in
resistance is linearly related to the change in temperature. The platinum wire
resistance thermometer can measure a very small change in temperature up
to +/-0.0001°C. The main disadvantage is slow response time.
The infrared thermometer absorbs infrared radiation emitted by the body
and converts the infrared signal into an electrical signal.
The mercury thermometer is not as accurate as the electronic methods.
Thermistors are made of semiconductor beads and contain a Wheatstone
bridge circuit. The resistance of the thermistor decreases non-linearly with
increasing temperature. They accurately measure the temperature to an
order of 0.1°C.
The Bourdon gauge thermometer is relatively simple, robust and cheap,
but not very accurate.