You are planning to administer a volatile anaesthetic using a completely closed breathing system where all the exhaled gases are re-breathed after absorption of CO2. Which of the following monitoring is the most essential?

18 views
  • Monitoring inspired oxygen concentration using a fuel cell at the common gas outlet.
  • Monitoring inspired oxygen concentration very close to the endotracheal tube. (Correct Answer)
  • Monitoring inspired CO2 concentration to detect hypercapnia.
  • Monitoring inspired concentration of volatile anaesthetic agent at the common gas outlet.

Explanation

Answer: B. Monitoring inspired oxygen
concentration very close to the endotracheal
tube.
In a completely closed system all the exhaled gases are rebreathed except for CO2 which is absorbed by the sodalime or baralime. The oxygen concentration in the exhaled gas mixture depends on the inspired oxygen concentration and alveolar oxygen extraction. The initial concentration of oxygen measured at the common gas outlet and anaesthetic agent concentration (dial setting) in the fresh gas flow are diluted in the circle system. The oxygen concentration gradually decreases over time due to the dilution effect from exhaled gases in the circle system. Therefore, it is essential to monitor the inspired oxygen concentration close to the endotracheal tube. Paramagnetic type oxygen analysers allow breath-to-
breath measurement of oxygen concentration.
Monitoring of inspired CO2 concentration is useful in detecting an exhausted CO2 absorber (sodalime).
Further reading
1. Davey AJ. Breathing systems and their components. In: Ward’s anaesthetic equipment, 5th ed. Davey AJ, Diba A, Eds. Philadelphia, USA: Elsevier Saunders, 2005; Chapter 7: 131-63.

Comments (0)

No comments yet. Be the first to comment!

Want to practice more?

Explore thousands of MCQs across every exam category on Exams Nepal.