You are performing an interscalene nerve block using a peripheral nerve locator on a 25-year-old woman with a BMI of 23. Which of the following is the most suitable needle for this block?
- A 22G, 50mm long insulated needle.
- A 20G, 100mm long insulated needle.
- A 22G, 25mm long non-insulated needle with a non-cutting tip.
- A 22G, 25mm long insulated needle with a non-cutting tip. (Correct Answer)
Explanation
Answer: D. 22G, 25mm long insulated needle
with a non-cutting tip. Purpose designed thin insulated needles are used in conjunction with the
nerve locator to precisely locate the nerve. When a non-insulated needle is used, the current disperses in all directions and, hence, a larger current
is needed to stimulate the nerve. In this patient (with a BMI of 23), the brachial plexus is superficial in the interscalene groove. Therefore, a needle longer than 25mm is unnecessary. A non-cutting needle reduces
the chance of nerve damage. A larger diameter needle can increase the risk of tissue damage. Hence, a 22G, 25mm insulated needle is preferred in this scenario.
Further reading
1. Dalrymple P, Chelliah S. Electrical nerve locators. British Journal of
Anaesthesia CEACCP 2006; 6: 32-6.