You are anesthetising a patient for magnetic resonance imaging (MRI). MRI can interfere with the monitoring equipment. Which one of the following best describes the problem of monitoring EtCO2 using side-stream capnography?
- The EtCO2 reading is higher than the true value
- The rise time increases.
- The transit time increases. (Correct Answer)
- The EtCO2 reading is lower than the true value.
Explanation
Correct Answer: C. The transit time increases.
The most significant problem associated with magnetic resonance imaging
(MRI) is the attraction of ferromagnetic objects to the magnetic field. The
displays on standard monitors can be distorted by the magnetic field and
the monitors themselves can degrade the MRI image by altering the
signal/noise ratio.
The transit time is the time required for the sample to move from the point
of sampling to the detector cell. In the MRI suite the side-stream analyser
should be kept away from the magnetic field. Therefore, a long sampling
tube is required. This prolongs the transit time. A prolonged transit time
delays the appearance of the waveform at the monitor resulting in phase
shift but no distortion. The actual EtCO2 reading is not affected.
The rise time is the time taken by the output from the capnometer to
change from 10% of final value to the 90% final value in response to a
step change in partial pressure of CO2. It is dependent on the size of the
sample chamber and the gas flow rate.
The response time is the combination of rise time and transit time. A
response time less than one respiratory cycle is ideal.