You are anaesthetising a patient at high altitude where atmospheric pressure is 380mm Hg. If the set concentration on the isoflurane vaporiser is 2%, in reality which one of the following indicates the correct concentration of vapour delivered in the inspiratory flow to the patient?

17 views
  • 1%.
  • 2%.
  • 3%.
  • 4%. (Correct Answer)

Explanation

Answer: D. 4%.
As the atmospheric pressure is reduced, the delivered concentration is increased from that marked on the dial of a vaporiser. Since the barometric pressure is reduced by half of that at sea level, the concentration of vapouroutput doubles. However, anaesthetic action depends on the alveolar partial pressure, and not on concentration. The partial pressure of isoflurane delivered would be approximately the same at both altitudes since 2% isoflurane at 760mm Hg (15.2mm Hg) is the same as 4% isoflurane at 380mm Hg (15.2mm Hg). Saturated vapour pressure (SVP)
is unaffected by atmospheric pressure and therefore the partial pressure of isoflurane delivered is the same as at sea level.
Further reading
1. Carter JA. Provision of anaesthesia in difficult situations and the developing world. In: Ward’s anaesthetic equipment, 5th ed. Davey
AJ, Diba A, Eds. Philadelphia, USA: Elsevier Saunders, 2005; Chapter 29: 485-98.