Which preventive step reduces the risk of inadvertent thermal injury from monopolar electrosurgery during laparoscopy?
27 views
- Use nonconductive (plastic) trocars and ensure safe instrument distances (Correct Answer)
- Always use maximum power to finish quickly
- Rely on wet gauze to conduct current away from tissues
- Keep instruments in contact with metal trocars at all times
Explanation
Nonconductive trocars interrupt current paths that lead to capacitive coupling or insulation failure burns; combined with attention to instrument activation, insulation checks and safe distances, this lowers thermal injury risk.