Which one of the following most accurately represents the precautions taken to minimise the risk of fire and explosion in the theatre environment?
- The relative humidity in theatre should be at least 85%.
- Substitution of cotton fabric with nylon.
- Use of only CO2 as insufflation gas during laparoscopic surgery. (Correct Answer)
- Maintaining a 10cm zone of risk around the anaesthetic machine.
Explanation
Answer: C. Use of only CO2 as insufflation gas
during laparoscopic surgery.
Precautions to be taken to minimize the risk of fire and explosions in the operating theatre include:
Avoidance of use of flammable agents in the operating room.
Use of spark-free switches.
Adequate air conditioning and scavenging with 15-20 changes of air per hour in the operating theatre.
Maintaining a relative humidity of more than 50% in the operating theatre. A dry atmosphere promotes the generation of static electrical charges. However, working in an atmosphere with very high humidity
(>80%) is quite uncomfortable.
Preferable use of circle systems.
Ensuring that flammable skin preparation solutions have evaporated
completely before using the diathermy.
Avoidance of build-up of static electricity.
The components of a fire triangle include an ignition source, fuel and an oxidizing agent to support combustion.
Static electricity is a known source of ignition in theatre. Materials that are likely to cause static charges such as nylon and wool should be avoided
in the theatre environment. Bowel gas contains methane (up to 30%) and hydrogen (up to 44%) which are highly flammable. Liberation of these gases from accidental perforation of the bowel could potentially cause ignition in the abdomen during surgery. Only 100% CO2 should be used for gas insufflation during laparoscopy. Although the oxygen concentration
in the intestines is low (less than 5%) and cannot support a fire, diffusion of nitrous oxide into the bowel and peritoneal cavity poses a potential danger.
The zone of risk is the area of the operating room in which mixtures of anaesthetic agents may be explosive. It extends to 25cm from any part of
the anaesthetic apparatus or the patient’s airway containing the anaesthetic mixture. Any flame or potential source of risk must be outside
this zone.