Which of the following best describes the intra-arterial cannula used for the purpose of invasive arterial blood pressure measurement?
- It should be short, wide, stiff and with parallel walls. (Correct Answer)
- It should be short, wide, stiff and tapered at the distal end.
- It should be short, wide, stiff and transparent.
- It should be made of polyurethane.
Explanation
Answer: A. It should be short, wide, stiff and with
parallel walls.
The diameter of the cannula used to cannulate the artery is a balance between a small cannula (22 or 20 gauge) that carries a lower incidence of thrombus formation and a larger catheter (18 or 16 gauge) that is less likely to kink or become blocked. Arterial cannulae are made of teflon or polyurethane and have parallel walls to minimize the effect on blood flow to the distal part of the limb. A short, wide, stiff and parallel-sided catheter is useful in minimising the effects on the resonant frequency of the system.
Further reading
1. Al-Shaikh B, Stacey S, Eds. Invasive monitoring. In: Essentials of
anaesthetic equipment, 3rd ed. London, UK: Churchill Livingstone,
Elsevier, 2007; Chapter 11: 162-3.
2. Davis PD, Kenny GNC. Blood pressure measurement. In: Basic
physics and measurement in anaesthesia, 5th ed. London, UK:
Butterworth-Heinemann, 2003; Chapter 17: 192-3.