Which of the following best describes the intra-arterial cannula used for the purpose of invasive arterial blood pressure measurement?
19 views
- It should be short, wide, stiff and with parallel walls. (Correct Answer)
- It should be short, wide, stiff and tapered at the distal end.
- It should be short, wide, stiff and transparent.
- It should be made of polyurethane
Explanation
Correct Answer: A. It should be short, wide, stiff and with
parallel walls.
The diameter of the cannula used to cannulate the artery is a balance
between a small cannula (22 or 20 gauge) that carries a lower incidence
of thrombus formation and a larger catheter (18 or 16 gauge) that is less
likely to kink or become blocked. Arterial cannulae are made of teflon or
polyurethane and have parallel walls to minimize the effect on blood flow
to the distal part of the limb. A short, wide, stiff and parallel-sided catheter
is useful in minimising the effects on the resonant frequency of the system.