Which lung volume or capacity CANNOT be measured by simple spirometry?

13 views
  • Tidal volume
  • Vital capacity
  • Inspiratory reserve volume
  • Functional residual capacity (Correct Answer)

Explanation

Spirometry measures volumes of air moved into and out of the lungs, so it can measure tidal volume (about 500 mL), inspiratory reserve volume (about 3000 mL), expiratory reserve volume (about 1100-1200 mL), vital capacity (about 4600-4800 mL), inspiratory capacity, FEV1 and FVC. It cannot measure residual volume (RV, about 1200 mL), the air remaining in the lungs after maximal expiration, so any capacity that includes RV, namely functional residual capacity (FRC = ERV + RV, about 2300-2400 mL) and total lung capacity (TLC = VC + RV, about 5800-6000 mL), also cannot be measured by spirometry. These are measured by helium dilution (closed-circuit), nitrogen washout (open-circuit, multiple-breath), or body plethysmography. Plethysmography (body box, based on Boyle's law) measures all gas in the thorax, including gas trapped behind closed airways, so it gives higher values than gas dilution methods in obstructive disease such as emphysema. FRC is the resting volume where the inward elastic recoil of the lung balances the outward recoil of the chest wall; it decreases supine, under anaesthesia, in obesity and in pregnancy, and increases in emphysema.