Which biological indicator is used to verify the efficacy of steam sterilisation in an autoclave?
- Bacillus atrophaeus spores
- Geobacillus stearothermophilus spores (Correct Answer)
- Clostridium tetani spores
- Bacillus pumilus spores
Explanation
Autoclaving uses saturated steam under pressure, killing microorganisms, including spores, by moist heat, which coagulates and denatures proteins more efficiently than dry heat. Standard cycles are 121 degrees C at 15 psi (about 1.1 kg/cm2) above atmospheric pressure for 15-20 minutes, or 134 degrees C for 3-4 minutes (used in flash and prion-related protocols; prions require 134 degrees C for 18 minutes or other special methods). Monitoring of sterilisation uses physical indicators (time, temperature and pressure records), chemical indicators (autoclave tape, Bowie-Dick test for air removal in pre-vacuum autoclaves), and biological indicators, which are the definitive test. The biological indicator for steam autoclaves (and also hydrogen peroxide plasma sterilisation) is Geobacillus (formerly Bacillus) stearothermophilus spores, which are highly heat resistant and thermophilic (growing at 55-60 degrees C). Bacillus atrophaeus (formerly B. subtilis var. niger) spores are used for dry heat (hot air oven) and ethylene oxide sterilisation. Bacillus pumilus was used for ionising radiation. Clostridium tetani spores were historically used to test hot air ovens (non-toxigenic strains). Correct loading, removal of air, cleaning before sterilisation and wrapping are as important as cycle parameters for infection prevention.