Under the Spaulding classification, a dental handpiece that contacts oral mucous membranes but does not penetrate tissue should be reprocessed by:
- Wiping with a low-level disinfectant between patients
- Heat sterilisation after each patient, since heat-tolerant semi-critical items should be sterilised (Correct Answer)
- High-level chemical disinfection only, with no cleaning required
- Cleaning with soap and water only, since it is a non-critical item
Explanation
The Spaulding classification (1968) groups patient care items by the risk of infection associated with their use. Critical items penetrate soft tissue or bone or enter sterile tissue or the vascular system. Examples in dentistry are surgical instruments, scalpels, burs used in surgery, periodontal scalers and endodontic files. They must be sterilised, usually by steam autoclave, or be single-use. Semi-critical items contact mucous membranes or non-intact skin but do not penetrate soft tissue or bone. Examples include mouth mirrors, amalgam condensers, impression trays and dental handpieces. The CDC guidelines state that heat-tolerant semi-critical items should be heat-sterilised after each use. High-level disinfection is acceptable only for heat-sensitive semi-critical items, and single-use disposable versions are preferred. Dental handpieces, including high-speed and low-speed handpieces, motor-driven attachments, air-water syringe tips and ultrasonic scaler tips, must be cleaned, lubricated and heat-sterilised between patients, because internal parts can be contaminated by aspiration of oral fluids. Surface disinfection alone is inadequate. Non-critical items contact only intact skin, for example blood pressure cuffs and radiograph tube heads. They need cleaning and low- or intermediate-level disinfection, and clinical contact surfaces can be protected with barriers changed between patients. All instruments must be cleaned first (manual, ultrasonic or washer-disinfector) because organic debris protects microorganisms from sterilisation.