Third-generation apex locators such as Root ZX determine the apical constriction by measuring:
- An absolute resistance of 6.5 kOhm
- The ratio of impedances at two different frequencies (Correct Answer)
- Capacitance change at a single frequency
- Change in direct current with pulsed signals
Explanation
Sunada (1962) showed that the electrical resistance between the oral mucosa and the periodontal ligament is relatively constant, about 6.5 kOhm. First-generation (resistance-based) apex locators used this principle, but conductive irrigants, pulp tissue and blood made them unreliable. Second-generation devices measured impedance at a single frequency and were also affected by canal contents. Third-generation devices use the frequency-dependent impedance method. Root ZX, developed by Kobayashi and Suda, uses the "ratio method": it measures impedance at two frequencies (8 kHz and 0.4 kHz) and calculates their ratio. The ratio stays fairly constant in the canal and changes sharply as the file tip approaches the apical constriction. Because the ratio is relatively unaffected by electrolytes, the device works in canals filled with NaOCl, EDTA or saline, although very wet canals can still cause premature readings. Fourth-generation devices measure resistance and capacitance separately, and later models use multiple frequencies. Accuracy in locating the minor diameter is roughly 80 to 90% within 0.5 mm, so radiographic confirmation is still recommended.