The Root ZX electronic apex locator determines the apical constriction by measuring:
- Direct current resistance
- Impedance at a single frequency
- The ratio of impedances at two frequencies (Correct Answer)
- Voltage gradient
Explanation
Electronic apex locators work on the principle that the electrical resistance between the periodontal ligament and the oral mucosa is roughly constant (about 6.5 kilohms, as shown by Suzuki and Sunada). First-generation locators measured direct current resistance and were inaccurate in wet canals. Second-generation devices used impedance at a single frequency but were still affected by electrolytes and required a dry canal. Third-generation devices such as the Root ZX (Kobayashi, 1994) use the ratio method: they measure impedance at two frequencies (8 kHz and 0.4 kHz) simultaneously and compute their ratio. The ratio stays fairly constant in the canal but decreases sharply as the file approaches the apical constriction. Because it is a ratio, it is largely unaffected by the type of irrigant, pulp tissue or blood in the canal, and it is highly accurate (about 90%). Readings can still be affected by open apices, root perforations, metal restorations and saliva contact.