The Pethick test is used to check:
- Leaks in the unidirectional valves of a circle system
- Accuracy of the oxygen analyser
- Function of the APL valve
- Integrity of the inner tube of a Bain circuit (Correct Answer)
Explanation
The Bain circuit is a coaxial modification of the Mapleson D system: fresh gas flows through the narrow INNER tube to the patient end, and exhaled gas passes through the wider OUTER tube towards the reservoir bag and APL valve. If the inner tube is disconnected, kinked or cracked at the machine end, fresh gas enters the outer tube directly. The whole outer limb then becomes dead space, causing massive rebreathing, hypercapnia and hypoxia, which can be missed on visual inspection. Pethick test: occlude the patient end, close the APL valve and fill the reservoir bag; then release the occlusion and activate the oxygen flush. The high-velocity flow at the patient end produces a Venturi effect that draws gas out of the outer tube, so the reservoir bag COLLAPSES if the inner tube is intact. If the inner tube is faulty, the bag inflates or fails to deflate. An alternative test is occluding the inner tube at the patient end with a syringe plunger, which should make the flowmeter bobbin dip. Recommended fresh gas flow for the Bain circuit is about 150-200 mL/kg/min for spontaneous ventilation and about 70-100 mL/kg/min for controlled ventilation. Mapleson A is the most efficient circuit for spontaneous ventilation.