The graph below describes the relationship between true blood pressure and measured blood pressure through an invasive arterial cannula connected to a transducer system (Figure 1). Line A represents the ideal response. Lines B and C represent false readings due to a calibration error. The error represented by line C can best be corrected by:

19 views
  • Zeroing the system.
  • Replacing the transducer cable.
  • Performing a three-point calibration. (Correct Answer)
  • Performing a two-point calibration.

Explanation

Answer: C. Performing a three-point calibration.
In order to obtain an accurate blood pressure reading, the system should be appropriately calibrated. Zero calibration eliminates the effect of atmospheric pressure on the measured pressure. Zeroing ensures that the monitor indicates zero pressure in the absence of applied pressure; it eliminates the offset drift (zero drift). To eliminate the gradient drift,
calibration at a higher pressure is necessary. For applying a known higher pressure, the transducer is connected to an aneroid manometer using
sterile tubing through a three-way stopcock and the manometer pressure is raised to 100 and 200mm Hg. The monitor display should read the same pressure as that applied to the transducer.
For accurate measurement the transducer system must be ‘zeroed’ to a reference point. This reference point is usually on the left side in the mid-axillary line at the level of the left ventricle.
Referencing or levelling the transducer system is accomplished by aligning the air-fluid interface of the transducer system (the three-way stopcock at the top of the transducer) to the mid-axillary point. Zeroing is then performed by opening the three-way stopcock between the patient and the transducer to atmosphere and selecting the zero on the monitor.
In the supine position, the mid-axillary line is an appropriate reference point. Raising or lowering the transducer above or below this point will result in an error equivalent to 7.5mm Hg for each 10cm change in the height. However, if the clinician is interested in measuring the MAP at the level of the brain in the sitting position, then a different reference point should be chosen.
Generation of a square wave at the catheter tip is the gold standard laboratory test in assessing the dynamic response of a monitoring system. It is used for assessing the optimum damping coefficient.
Further reading
1. Davis PD, Kenny GNC. Presentation and handling of data and basic measurement concepts. Basic physics and measurement in anaesthesia, 5th ed. London, UK: Butterworth-Heinemann, 2003; Chapter 25: 285-8.
2. Kleinman B, Powell S, Gardner RM. Equivalence of fast flush and square wave testing of blood pressure monitoring systems. Journal of Clinical Monitoring 1996; 12: 149-54.

Comments (0)

No comments yet. Be the first to comment!

Want to practice more?

Explore thousands of MCQs across every exam category on Exams Nepal.

Find More MCQs