The following are true regarding sources of error in pulse oximetry EXCEPT:
- Foetal haemoglobin (HbF) causes overestimation of SpO2 (Correct Answer)
- Jaundice does not affect the signal.
- Severe tricuspid regurgitation reduces the SpO2 reading
- Readings are unreliable below 70% SpO2.
Explanation
Correct Answer: A.
Prilocaine is a cause of methaemoglobinaemia, which tends to cause the
SpO2 to read around 85%. Jaundice, foetal haemoglobin and dark skin do not affect the accuracy of the measurement. The pulse oximeter estimates SpO2 from the pulsatile component of the measured absorption spectrum (discarding the non-pulsatile venous and tissue component). Severe tricuspid regurgitation causes venous pulsation in the tissues (especially with ear probes), which causes deoxygenated but pulsatile venous blood to be measured, falsely lowering SpO2. Readings below 70% are extrapolated rather than from direct reference data, and are therefore unreliable. However, any reading below 90% indicates serious hypoxaemia due to the steep fall in the oxygen dissociation curve at this point.