The figure below is an arterial trace from a 70-year-old patient with chronic obstructive airway disease, in the intensive care unit. This trace indicates:
- Presence of blood clot in the cannula.
- An under-damped trace (Correct Answer)
- Compliant tubing.
- Atrial fibrillation.
Explanation
Answer: B. An under-damped trace.
The arterial trace presented in the question shows a falsely high systolic
pressure and a falsely low diastolic pressure. The mean arterial pressure
is unaffected. An under-damped arterial trace is recognised by the
presence of an overshoot spike (ringing). Increased resonance can be due
to a stiff, non-compliant diaphragm and tubing. The waveform is underdamped.
Over-damping (damping) results in a smoothed out trace without
displaying sharp changes, leading to under-reading of systolic pressure
and over-reading of diastolic pressure. The loss of pressure in the fluidfilled tubing system, soft compliant tubing, numerous connections and
stopcocks, kinking of the cannula, blood clots and air bubbles can result
in an over-damped arterial trace.