The biological indicator used to monitor the efficacy of steam autoclaving is spores of:

13 views
  • Bacillus atrophaeus
  • Geobacillus stearothermophilus (Correct Answer)
  • Bacillus pumilus
  • Clostridium tetani

Explanation

Geobacillus stearothermophilus (formerly Bacillus stearothermophilus) is a thermophilic spore-forming bacterium whose spores are highly resistant to moist heat, making it the standard biological indicator for autoclaves (121°C at 15 lb/sq inch for 15 minutes, or 134°C for 3 minutes). The spore strips are placed in the load, autoclaved, and then incubated at 55-60°C for up to 7 days; absence of growth confirms successful sterilization. It is also used for hydrogen peroxide plasma sterilization. Bacillus atrophaeus (formerly B. subtilis var. niger/globigii) is the indicator for dry heat (hot air oven) and ethylene oxide sterilization. Bacillus pumilus is used for ionizing radiation. Clostridium tetani spores were historically used for hot air oven testing (non-toxigenic strain) but are not used for autoclaves. In dental practice, spore testing of autoclaves is recommended at least weekly, while chemical indicators (e.g. Bowie-Dick test) are used for each cycle to check air removal and steam penetration.