The biological indicator used to monitor steam autoclave sterilization is spores of:
- Bacillus atrophaeus
- Geobacillus stearothermophilus (Correct Answer)
- Clostridium tetani
- Bacillus pumilus
Explanation
Biological indicators use highly resistant bacterial spores and are the most reliable way to confirm that sterilization conditions were achieved. Steam autoclave and chemical vapour (unsaturated chemical vapour) sterilization are monitored with Geobacillus (formerly Bacillus) stearothermophilus, which is thermophilic and highly resistant to moist heat; after the cycle, the strip is incubated at 55-60 degrees C. Dry heat and ethylene oxide sterilization are monitored with Bacillus atrophaeus (formerly B. subtilis var. niger). Ionizing radiation is monitored with Bacillus pumilus. Hydrogen peroxide plasma uses G. stearothermophilus. CDC guidelines recommend at least weekly spore testing of dental sterilizers, plus testing with every load containing an implantable device and after repair. Mechanical indicators (time, temperature, pressure readouts) and chemical indicators (external and internal, e.g., class 5 integrators, and Bowie-Dick test for air removal in prevacuum autoclaves) are used for every cycle. Standard autoclave cycles: 121 degrees C at 15 psi for 15-20 minutes (gravity) or 134 degrees C at 30 psi for 3-4 minutes (prevacuum). Clostridium tetani is not a standard biological indicator.