Regular maintenance of electrodes in blood gas analyzers is essential. In the CO2 electrode, which of the following components most needs to be replaced at regular intervals?
- The membrane covering the electrode. (Correct Answer)
- The electrode itself.
- The electrolyte solution.
- The pump tubings
Explanation
Correct Answer: A. The membrane covering the electrode.
In the CO2 electrode, CO2 diffuses through the membrane and reacts with
water in the electrolyte solution to produce hydrogen ions. The hydrogen
ion concentration is measured with a pH-sensitive glass electrode. As with
other electrodes, the CO2 electrode needs regular maintenance. Integrity
of the membrane is essential for the accuracy of the electrode. If a hole
occurs in the membrane it ceases to be a semi-permeable membrane.
The CO2 electrode is calibrated using two different gases with known
concentrations of CO2. The electrode itself does not require regular
replacement, but other components such as pump tubing, calibrating
solution and rinse fluid should be replaced at regular intervals.
In the automated modern blood gas analyser, by selecting the analyser
status screen on the display monitor, the time for electrode and membrane
replacement can be checked.