On a Levey-Jennings chart, two consecutive control results for the same level both fall between +2 SD and +3 SD. According to Westgard rules, this indicates:
- A 1-2s warning only, which can be ignored if other controls are acceptable
- A 2-2s violation suggesting systematic error, so the run should be rejected (Correct Answer)
- An R-4s violation indicating random error
- A 10x violation indicating a long-term trend
Explanation
The 2-2s rule is violated when two consecutive control results exceed the same 2 SD limit on the same side of the mean, either within one control level across consecutive runs or across two control levels within a single run. It indicates systematic error, such as deterioration of a reagent, calibration drift, a change of lamp or a temperature problem, and the run should be rejected while the cause is investigated and corrected before patient results are released. A 1-2s result is a warning that triggers inspection of other rules, but here the 2-2s rule is clearly broken. The R-4s rule applies when one control exceeds +2 SD and another exceeds -2 SD within the same run, a range greater than 4 SD, which indicates random error such as pipetting problems, bubbles or sample mix-ups. The 10x rule requires ten consecutive results on the same side of the mean. Correctly identifying the type of error directs troubleshooting toward the most likely cause.