Medical monitoring equipment is classified according to the means of protection it provides against electric shock. Which of the following statements best describes class 2 equipment?

14 views
  • Any conducting part that is accessible to the user is connected to the earth without a fuse in the mains plug.
  • It is double insulated equipment and is connected to the earth with a fuse in the main plug.
  • It is double insulated equipment and the earth wire is not essential (Correct Answer)
  • It is internally powered equipment with minimal risk of electric shock

Explanation

Answer: C. It is double insulated equipment and the earth wire is not essential.

To ensure safety and prevent electrical hazards, all medical equipment
should meet the requirement of certain national and international
standards. IEC (international electronic commission) 60601-1, introduced
in 2005, describes the general requirements for basic safety and essential
performance of medical equipment.
Medical monitoring equipment is classified according to the means of
protection against electrical shock:
w Class 1: any conducting part of the equipment, which may contact the
patient, is connected to earth by an earth wire. It incorporates a fuse
in the mains plug that melts to break the circuit when a live supply
comes into contact with the accessible part, and it also has a fuse in
the live and neutral conductors for additional protection.
w Class 2: this has a double insulation or reinforced insulation. An earth
wire is not required.
w Class 3: this is battery-powered equipment with a voltage not
exceeding 25V AC or 60V DC. Even with this low voltage, the risk of
microshock still exists.