For spontaneous ventilation, the Mapleson system A (Magill’s circuit) is the most efficient system. What is the single most important reason for this?
19 views
- Requires low fresh gas flow
- Does not allow wastage of fresh gas flow.
- Dead space gas is re-used. (Correct Answer)
- Fresh gas flow is equal to alveolar ventilation.
Explanation
Correct Answer: C. Dead space gas is re-used.
Although all statements are correct, the most important reason is that the
dead space gas is re-used. In a Mapelson A system, the spill valve
(adjustable pressure relief valve) is located at the patient end and the
reservoir bag at the machine end. During the initial part of expiration the
dead space gas, which does not contain CO2, flows backwards towards
the reservoir bag. The reservoir bag continues to fill with fresh gas flow.
Once it is full the spill valve opens and alveolar gas containing CO2 from
the patient is exhaled.