For intraoral radiography, the maximum permissible diameter of a circular X-ray beam at the patient's skin surface is:

32 views
  • 7 cm (2.75 inches) (Correct Answer)
  • 10 cm (4 inches)
  • 12 cm (4.7 inches)
  • 5 cm (2 inches)

Explanation

Collimation restricts the size and shape of the X-ray beam to reduce the volume of tissue irradiated and to decrease scatter, which improves image contrast. International and national guidance, including NCRP recommendations followed in most dental radiology texts, limits a circular intraoral beam to a diameter of no more than 7 cm (2.75 inches) at the patient's face. Rectangular collimation, sized slightly larger than a size 2 receptor (about 3.5 x 4.5 cm), is strongly recommended because it reduces the irradiated area, and therefore patient dose, substantially compared with a 7 cm round beam; it is now recommended as standard practice. To limit the risk of cone cutting with rectangular collimation, film holders with aiming rings (beam-alignment devices) should be used. Lead-lined position-indicating devices (PIDs) that are long and open-ended are preferred over short pointed cones, which were discarded because they produced more scatter. Other dose-reduction measures include using the fastest image receptor (F-speed film or digital sensor), adequate filtration (at least 1.5 mm aluminium up to 70 kVp, 2.5 mm above 70 kVp), thyroid collars where they do not obscure the area of interest, and selection criteria so that radiographs are taken only when clinically justified.

Comments (0)

No comments yet. Be the first to comment!