Fixed dose oxygen can be delivered by:
- Nasal prongs
- Face mask
- Venturi mask (Correct Answer)
- Laryngeal mask
Explanation
Ans c) Venturi mask
# High Flow /Fixed Performance Equipments
- O2 is supplied at predetermined rate & at fixed concentration inspite of patients respiratory
pattern (i.e flow rate is not affected by respiratory pattern). This is achieved by delivering O2 at a flow rate greater than the peak inspiratory flow rate. A flow of more than 20L/min is adequate because a normal inspiratory flow rate is 20-30L/ min for an, adult. Because of high flow rate expired gases are rapidly flushed out and thus, there is no rebreathing & no increase in dead space.
- Venturi mask is colour coded and marked with recommended oxygen flow rate. Maximum O2 that can be delivered by venture mask is 60%. Indications of ventimask (fixed performance device are consistent Fi02 & /or large inspiratory flow of gas (>40L/min).
# Low flow or variable performance Equipments
- Provides a fraction of inspired O2. Here the Fi02. The patient is getting is uncertain. It is only used when it is not critical to deliver a fixed oxygen concentration.
- The final conc. of Inspired O2 depends upon O2 supply flow rate, how tight the mask fits on the face, the pattern of ventilation, patients inspiratory flow rate.
- Low flow devices are nasal cannula (maximum 02 delivered is 44%), Mary-careterall mask, nasal mask, oxygen tents, simple mask, BLB mask rebreathing and non- rebreathing masks.