During low-flow anaesthesia, the inspired oxygen concentration in fresh gas flow can be further diluted. Which of the following is most useful in preventing the delivery of hypoxic gas mixtures during lowflow anaesthesia using a fresh gas flow of 0.3L/minute?
- Mechanical link 25 system.
- Pneupac ratio system.
- Penlon electronic system.
- Breath-to-breath oxygen monitoring close to the endotracheal tube. (Correct Answer)
Explanation
Correct Answer: D. Breath-to-breath oxygen monitoring
close to the endotracheal tube.
When oxygen and nitrous oxide are used, it is possible to deliver a hypoxic
gas mixture, if inspired oxygen concentration is not constantly monitored.
A mechanical link 25 system uses a chain to link flow control valves for
oxygen and nitrous oxide. This system may fail to account for other gases,
such as air, reducing the oxygen concentration to less than 25%.
A Pneupac ratio system uses a ratio mixer valve. Oxygen supplied to this
valve exerts a pressure on one side of the diaphragm, which is opposed
by the pressure of nitrous oxide on the opposite side. The diaphragm
construction ensures an increase in oxygen flow rate by a ratio of 25% of
any increase in the nitrous oxide flow rate.
Electronic devices (e.g. Penlon) use a paramagnetic oxygen analyser to
continuously sample the gas mixtures from the flow meters. If the inspired
oxygen fraction decreases below 0.25, the nitrous oxide is temporarily cut
off, whereas an increase in the inspired oxygen fraction will temporarily
restore nitrous oxide flow.
All of the above three systems ensure that a minimum of 25% oxygen is
delivered from the common gas outlet. However, during low-flow
anaesthesia, other gases and vapours dilute the inspired oxygen
concentration in the breathing system. Therefore, a system that ensures
25% oxygen concentration at the common gas outlet is not sufficient in
preventing the delivery of a hypoxic mixture to the patient. The inspired
oxygen concentration needs to be monitored close to the endotracheal
tube to avoid delivering hypoxic gas mixtures.
The Ritchie whistle is powered by an oxygen supply at a pressure of
420kPa. At a pressure below 200kPa it cuts off the supply of anaesthetic
gases and allows the patient to inspire room air. The whistle sounds
continuously until the pressure has fallen to 40.5kPa.