According to the Spaulding classification, a flexible fibreoptic gastroscope that contacts intact mucous membranes but not sterile tissue requires which minimum level of reprocessing?
- Sterilisation by steam autoclave at 134 degrees Celsius
- High-level disinfection, for example with 2% glutaraldehyde or ortho-phthalaldehyde (Correct Answer)
- Cleaning with detergent and water only
- Low-level disinfection with 70% alcohol wipe only
Explanation
The SPAULDING CLASSIFICATION assigns reprocessing requirements for medical devices based on the RISK OF INFECTION from their use. (1) CRITICAL items enter STERILE tissue or the vascular system: surgical instruments, implants, urinary catheters, needles, cardiac catheters and arthroscopes or laparoscopes. They require STERILISATION (destruction of ALL microorganisms, including bacterial SPORES): steam autoclave (121 degrees Celsius for 15-20 minutes at 15 psi, or 134 degrees Celsius for 3-4 minutes), dry heat, ethylene oxide gas, hydrogen peroxide plasma, or prolonged immersion in a liquid chemical sterilant for heat-sensitive items. (2) SEMI-CRITICAL items contact MUCOUS MEMBRANES or NON-INTACT skin: FLEXIBLE ENDOSCOPES (gastroscopes, colonoscopes, bronchoscopes), laryngoscope blades, endotracheal tubes, anaesthesia and respiratory therapy equipment, vaginal ultrasound probes and diaphragm fitting rings. They require at least HIGH-LEVEL DISINFECTION (HLD), which kills all vegetative bacteria, mycobacteria, fungi and viruses, but not necessarily large numbers of spores. Agents: 2% GLUTARALDEHYDE (Cidex, 20 minutes or more for HLD; 10 hours for sterilisation), ORTHO-PHTHALALDEHYDE (OPA, faster and less irritant), hydrogen peroxide, peracetic acid, and pasteurisation. Intact mucosa resists common spores but is susceptible to mycobacteria and viruses. Flexible endoscopes are HEAT-SENSITIVE, so they cannot be autoclaved. (3) NON-CRITICAL items contact only INTACT SKIN: BP cuffs, stethoscopes, bedpans, bed rails, crutches and ECG leads. They need LOW-LEVEL disinfection (70% alcohol, quaternary ammonium compounds, dilute hypochlorite) or cleaning. ALWAYS perform thorough MECHANICAL CLEANING before disinfection or sterilisation, because organic matter (bioburden) inactivates disinfectants and shields organisms. Endoscope-transmitted outbreaks (for example carbapenem-resistant Enterobacteriaceae with duodenoscopes) usually follow cleaning failures.