A volumetric (bulb) pipette marked "TD" without a frosted band near the top is being used to deliver 10.0 mL of a standard solution. Which technique delivers the calibrated volume correctly?

32 views
  • Let it drain under gravity, touch the tip to the vessel wall, and leave the residual drop in the tip (Correct Answer)
  • Allow the liquid to drain freely and then blow out the residual drop in the tip with a pipette bulb
  • Expel the liquid rapidly with the pipette bulb to minimise drainage time and evaporation
  • Rinse the pipette with distilled water after delivery and add the rinsing to the receiving vessel

Explanation

A "TD" (to deliver) volumetric pipette is calibrated so that the stated volume is delivered when the liquid drains under gravity and the tip is touched against the wall of the receiving vessel; the small drop that remains in the tip is accounted for in the calibration and must not be expelled. Blowing out the last drop is correct only for blow-out pipettes, which are identified by a frosted or etched band near the top. Forcing the liquid out rapidly leaves a variable film on the inner wall and gives an inaccurate, irreproducible volume. Rinsing the pipette into the vessel is the technique for "TC" (to contain) pipettes, such as some micropipettes used for viscous samples, and would add extra volume with a TD pipette. Reading the meniscus at eye level, using clean grease-free glassware and respecting the drainage time are other basic requirements for accurate volumetric work in the biochemistry laboratory.