A Tec 6 desflurane vaporizer is used where barometric pressure is 640 mmHg. To achieve the same anaesthetic partial pressure as a 6% setting at sea level, the dial should be set at approximately:

25 views
  • 5.0%
  • 6.0%
  • 7.1% (Correct Answer)
  • 9.0%

Explanation

Desflurane has a saturated vapour pressure of 669 mmHg at 20 degrees C and boils at 22.8 degrees C, so it cannot be used in a conventional variable-bypass vaporizer. The Tec 6 heats the agent to 39 degrees C and pressurises it to about 2 atm (around 1500 mmHg), then injects vapour into the fresh gas to deliver a constant VOLUME PERCENT regardless of ambient pressure. Anaesthetic effect, however, depends on PARTIAL PRESSURE. At sea level, 6% gives 0.06 x 760 = 45.6 mmHg. At 640 mmHg, 6% gives only 0.06 x 640 = 38.4 mmHg, which is under-dosing. Required setting = 45.6 / 640 = 0.071, or about 7.1%. Formula: required % = normal % x 760 / ambient pressure. In contrast, variable-bypass vaporizers (sevoflurane, isoflurane) deliver a constant partial pressure because SVP depends only on temperature; at altitude their volume % output rises automatically, so no correction is needed. This is practically relevant for hospitals at altitude in Nepal and other mountainous regions.