A patient with cardiogenic shock is having an intra-aortic balloon pump inserted. Which of the following will you rule out before using the pump?
- Acute myocardial infarction.
- Acute mitral regurgitation.
- Aortic regurgitation. (Correct Answer)
- Unstable angina.
Explanation
Answer: C. Aortic regurgitation.
An intra-aortic balloon pump (IABP) is a circulatory assist device in
critically ill patients with cardiac disease. The basic principle is counter
pulsation. The balloon is inflated in diastole and deflated in early systole.
Balloon inflation in diastole causes volume displacement of blood within
the aorta, both proximally and distally. This leads to a potential increase in coronary blood flow and systemic perfusion by augmentation of the
intrinsic Windkessel effect, whereby potential energy stored in the aortic
root during systole is converted to kinetic energy with the elastic recoil of
the aortic root.
Indications for IABP include the following:
* Acute myocardial infarction.
* Cardiogenic shock.
* Acute mitral regurgitation and ventricular septal defect.
* Catheterisation and angioplasty.
* Refractory unstable angina.
* Refractory ventricular arrhythmias.
* Refractory left ventricular failure.
* Sepsis.
Absolute contra-indications for IABP are:
* Aortic regurgitation.
* Aortic dissection.
* Aortic stents.
* Chronic end-stage heart disease with no anticipation of recovery.
Reference
1. Krishna M, Zacharowski M. Principles of intra-aortic balloon pump
counterpulsation. British Journal of Anaesthesia CEACCP 2009; 9:
24-8.