A patient is fitted with a VVI pacemaker. Which one of the following is INCORRECT regarding the functions of this pacemaker?

18 views
  • An intrinsic rhythm of a P wave transmitted to the ventricles will result in inhibition of pacing output
  • An ectopic of ventricular origin will result in inhibition of pacing output
  • A supraventricular tachycardia (SVT) transmitted to the ventricles will result in inhibition of pacing output
  • Administration of glycopyrronium bromide is more likely to increase than decrease the rate of pacing output (Correct Answer)

Explanation

Correct Answer: D. Administration of glycopyrronium bromide is more likely to increase than decrease the rate of pacing output

VVI pacemakers are inhibited by intrinsic ventricular activity. Polyfocalectopics
caused by wire insertion may result in significant compensatory pauses, thus resulting
in activation of pacing (if only temporarily). Glycopyrronium will increase sinoatrial
node activity, thus increasing intrinsic heart rate, and this will decrease rather than
increase the rate of pacing output.

Comments (0)

No comments yet. Be the first to comment!

Want to practice more?

Explore thousands of MCQs across every exam category on Exams Nepal.

Find More MCQs