A 76-year-old female patient with ischaemic heart disease is undergoing a laparotomy for a small bowel resection. Invasive arterial blood pressure is being monitored through a 20-gauge arterial cannula in the left radial artery. The transducer system is secured on a drip stand and zeroed with reference to the mid-axillary point. On request from the surgeon, the table is lowered by 10cm and the recorded mean arterial pressure (MAP) is 70mmHg. Which of the following is the true blood pressure at this time?
- 75mmHg.
- 80mmHg.
- 60mmHg.
- 77.6mmHg. (Correct Answer)
Explanation
Answer: D. 77.6mmHg.
The transducer system is zeroed with the reference point of the transducer at the level of the aortic root to eliminate the effect of the fluid column of
the system on blood pressure readings. Once zeroed, the transducer should be maintained at the same reference point. Since the table is lowered by 10cm, the transducer is now 10cm higher than the original
reference point. Therefore, it underestimates blood pressure by 10cm H2O which is equivalent to 7.6mmHg.
Further reading
1. Bedford RF, Shah NK. Blood pressure monitoring. In: Monitoring in
anaesthesia and critical care, 3rd ed. Blitt CD, Hines RL, Eds. New
York, USA: Churchill Livingstone, 1995: 95-130.