A 68-year-old male patient is undergoing a total knee replacement. The airway is secured with an i-Gel® supraglottic airway. Anaesthesia is maintained with sevoflurane 2.5% in oxygen and nitrous oxide with a total fresh gas flow of 0.8L/minute through a circle breathing system. The patient is breathing spontaneously. About an hour after starting the procedure, the patient showed signs of being in a light plane of anaesthesia, despite the sevoflurane dial set at 3%. The inspired sevoflurane is 2.6% and expired sevoflurane is 1.2%. Which of the following is the most likely cause for the gross difference in the inspired and expired concentration of sevoflurane?
- Malfunction of the vaporiser.
- Malfunction of the vapour analyser.
- Increased uptake of sevoflurane by the patient.
- Suboptimal positioning of the i-Gel® airway resulting in air entrainment. (Correct Answer)
Explanation
Answer: D. Suboptimal positioning of the i-Gel®
airway resulting in air entrainment.
Suboptimal insertion of an i-Gel® airway results in the gastric channel being open to the larynx. In a spontaneously breathing patient, air can be
entrained through the gastric channel which dilutes the anaesthetic gases. As this happens downstream to the point of agent monitoring (near the
catheter mount), the inspired concentration reading is not affected.
The patient has been anaesthetised for approximately one hour; equilibration between the alveolar and brain concentrations of sevoflurane should have occurred. Therefore, there should be no gross difference between inspired and expired concentrations despite low fresh gas flow.
Malfunctioning of the vaporiser would result in inaccurate vapour delivery. In this scenario the inspired concentration closely resembles the dial
setting. The use of low fresh gas flow is likely to result in a gross difference between the dial setting and inspired concentration measured by the agent
analyser.
Further reading
1. Intersurgical Ltd. Intersurgical i-Gel® User Guide, Issue 5.
Wokingham, UK: Intersurgical Ltd, 2008.
2. Baxter S. Phenomenon with i-gel airway Anaesthesia 2008; 63:1265.