A 67-year-old male patient is undergoing a total hip replacement under general anaesthesia. The airway is secured with a tracheal tube, and anaesthesia is maintained with sevoflurane (3%) and nitrous oxide (60%), and oxygen (40%) using a fresh gas flow of 0.3L/minute. Although the sevoflurane dial is set at 3%, the agent analyser indicates an inspired sevoflurane concentration of 1.8%. Which of the following is the most appropriate explanation for the difference between the dial setting and the inspired concentration?
- Malfunctioning vaporiser.
- Malfunctioning agent analyser.
- Ventilator malfunction.
- The dilutional effect of rebreathing. (Correct Answer)
Explanation
Correct Answer: D. The dilutional effect of rebreathing.
This is a common clinical observation during low-flow anaesthesia. The
dilutional effect of rebreathing contributes to the difference between the
dial setting and inspired concentration. The dial setting on the vaporiser
reflects the concentration delivered to the breathing system. The inspired
concentration detected by the agent analyser reflects the concentration of
anaesthetic agent at the tracheal tube end of the breathing system.
In low-flow anaesthesia, exhaled gases are recirculated after eliminating
the CO2 through the CO2 absorber (sodalime or baralime). The minute
volume is composed of the fresh gas flow (FGF) and the recirculated and
rebreathed exhaled gases. As FGF is decreased, the exhaled patient
gases contribute a more significant portion of the minute volume. In this
scenario, anaesthesia is maintained with a low FGF of 0.3L/minute; the
most likely reason for the difference is a dilution effect of the rebreathed
gases.
Malfunction of the vaporiser can deliver a lower concentration of
anesthetic and malfunction of the agent analyser can also account for the
inaccuracy in the measured concentration. Increased uptake of volatile
agent by the patient accounts for the increased difference between the
inspired and expired concentration rather than increased difference
between the dial setting and inspired concentration.