A 65-year-old man known to have COPD is admitted with severe respiratory distress. What is the flow rate required to deliver 28% oxygen through a venturi mask?
- 4L/minute (Correct Answer)
- 6L/minute
- 8L/minute
- 10L/minute
Explanation
Answer: A. 4L/minute.
The FiO2 delivered by venturi mask depends on the oxygen flow rate and the amount of air entrained. The amount of air entrained depends on the venturi mask adaptors. There are slits in the venturi mask adaptors, which become smaller or larger depending on whether a high or lower FIO2 is required. These slits are designed to entrain air from the environment.
Therefore, the set oxygen flow and the entrained air together meet the inspiratory flow requirement of the patient. The average FiO2 (fraction of inspired oxygen) delivered through venturi masks can be up to 5% above
the expected value.
Venturi masks provide a higher gas flow than the peak inspiratory flow rate irrespective of the patient’s respiratory pattern. The FiO2 delivered depends on the combination of venturi mask adaptor and the oxygen flow
rate. Therefore, just increasing the oxygen flow rate but keeping the adaptor constant does not increase the FiO2 delivered.
Venturi adaptors and oxygen flow rates are shown in Table 1.
Reference
1. Al-Shaikh B, Stacey S, Eds. Fixed performance devices. In:
Essentials of anaesthetic equipment, 3rd ed. London, UK: Churchill
Livingstone, Elsevier, 2007; Chapter 6: 89-91.