A 60-year-old male patient is undergoing a craniotomy and blood pressure is being invasively monitored via an arterial cannula inserted through the left radial artery. Half way through the procedure you wish to perform a square wave test (fast flush test) to ensure that the arterial system is optimally damped. Which of the following describes under-damping of the system?
- The waveform settles to zero without any oscillations.
- The waveform settles to zero after several oscillations. (Correct Answer)
- The waveform settles to zero after 2 or 3 oscillations.
- The waveform never settles to zero.
Explanation
Answer: B. The waveform settles to zero after
several oscillations.
To determine the optimum damping of the system a square wave test (fast
flush test) is used.
To perform the square wave test the system is flushed by applying a pressure of 300mm Hg (the flush button is compressed and released or the lever located near the transducer is pulled). This results in a square
waveform followed by oscillations.
In an optimally damped system, there will be two or three oscillations before settling to zero. An over-damped system settles to zero without any
oscillations. In an under-damped system, the waveform settles to zero after several oscillations.
The damping coefficient indicates how fast the oscillating system will come to rest.
If the damping coefficient = 0, then there is no damping, so oscillation will continue indefinitely.
If the damping coefficient =1 (critically damped), there is just enough damping to prevent oscillations.
For a fluid-filled catheter system, a damping coefficient = 0.67 is
considered as optimal damping.