A 50-year-old male patient is ready to be transferred from a district general hospital to a neurosurgical centre by ambulance. He is mechanically ventilated to achieve a minute volume of 5L/minute, using a portable ventilator, which consumes 5L/minute of oxygen to drive the ventilator. The total journey time is about 45 minutes. The ventilator requires 20 bar of pressure to operate. How many E-sized oxygen cylinders would be needed to complete the journey?
- One. (Correct Answer)
- Two.
- Three.
- Four.
Explanation
Answer: A. One.
The pressure in the full oxygen cylinder is 137 bar and the volume of the cylinder is 5L. A full size E oxygen cylinder contains 685L (137 x 5) of oxygen. But 5L will remain in the cylinder, so the available volume is 680L. Since the ventilator requires 20 bar of pressure to operate, it will stop operating when the cylinder pressure reaches 20 bar. Therefore, the available oxygen for clinical use is 580L (117 x 5 = 585-5). The total oxygen consumption is 10L per minute (5L is for ventilating the patient’s lungs and 5L for driving the ventilator). So the cylinder should last for 58 minutes. Therefore, one full cylinder of oxygen is required for the journey. However, in practical terms, one should ensure that double the amount of calculated oxygen is available to overcome any delays that may arise
during transfer.
Further reading
1. Davis PD, Kenny GNC. The gas laws. In: Basic physics and
measurement in anaesthesia, 5th ed. London, UK: Butterworth-
Heinemann, 2003: 37-50.
2. Bland H. The supply of anaesthetic and other medical gases. In:
Davey AJ, Diba Ali, Eds. Ward’s anaesthetic equipment, 5th ed.
Philadelphia, USA: Elsevier Saunders, 2005: 23-49.