A 4kg infant is admitted for a right inguinal herniotomy. He was born at 28 weeks’ gestation. Gas induction was discussed with the parents. What is the fresh gas flow (ml/minute) required to prevent rebreathing with a T-piece circuit?

20 views
  • 600ml/ minute..
  • 1000ml/minute..
  • 1600ml/minute.
  • 2400ml/minute. (Correct Answer)

Explanation

This child weighs 4kg. The tidal volume required is 10ml per kg of body weight. Minute volume is tidal volume x respiratory rate. In an infant, the
respiratory rate is around 20 breaths per minute.
A T-piece system requires a fresh gas flow of 2.5 to 3 times the minute volume to prevent rebreathing.
This child requires 40ml of tidal volume. So the minute volume is 800ml.
Minimum gas flow = 800 x 2.5-3 = 2000-2400ml/minute.
The T-piece is a valveless breathing system, particularly used in children weighing up to 25-30kg. It is suitable for both spontaneous and controlled
ventilation. It comprises a T-shaped tubing with three ports: the first port for fresh gas flow, the second port goes to the patient and the third port connects to the reservoir tubing. The system requires a fresh gas flow of 2.5 to 3 times the minute volume to prevent rebreathing with a minimal flow of 4L/minute. It requires a high fresh gas flow during spontaneous
ventilation. Since there is no APL valve in this breathing system, scavenging is a problem.
Further reading
1. Al-Shaikh B, Stacey S. T-piece system. In: Essentials of anaesthetic
equipment, 3rd ed. Parkinson M. London, UK: Churchill Livingstone,
Elsevier, 2007; Chapter 4: 53-4.

Comments (0)

No comments yet. Be the first to comment!

Want to practice more?

Explore thousands of MCQs across every exam category on Exams Nepal.

Find More MCQs