A 4kg infant is admitted for a right inguinal herniotomy. He was born at 28 weeks’ gestation. Gas induction was discussed with the parents. What is the fresh gas flow (ml/minute) required to prevent rebreathing with a T-piece circuit?

24 views
  • 600ml/ minute
  • 1000ml/minute
  • 1600ml/minute
  • 2400ml/minute (Correct Answer)

Explanation

Answer: D. 2400ml/minute. This child weighs 4kg. The tidal volume required is 10ml per kg of body weight. Minute volume is tidal volume x respiratory rate. In an infant, the respiratory rate is around 20 breaths per minute.
A T-piece system requires a fresh gas flow of 2.5 to 3 times the minute volume to prevent rebreathing