A 45-year-old male patient is undergoing a mastoidectomy under general anaesthesia. He is ventilated using a circle absorber system and low-flow anaesthesia. Which of the following is the single most important factor in preventing a critical incident related to a breathing system disconnection?
- Pulse oximeter.
- Capnography.
- Electronic airway pressure monitor.
- Vigilant anaesthetist continuously monitoring chest wall excursion. (Correct Answer)
Explanation
Answer: D. Vigilant anaesthetist continuously
monitoring chest wall excursion.
Monitoring devices supplement clinical observation. A vigilant anaesthetist is the vital link interpreting the monitoring and acting immediately to appropriately manage the critical incident. There are several disconnection monitors used in clinical practice. However, the most important monitor is
a vigilant anaesthetist who can interpret the physiological and mechanical (spirometry and pressure sensor) monitors. Disconnection of the
breathing system can be detected using the following measures.
Clinical observation:
- chest wall excursion;
- movement of ventilator bellows;
- absence of breath sounds.
Mechanical monitors:
- pneumotachograph;
- pressure sensors;
- capnography.
Monitoring EtCO2 is probably the best monitor in revealing patient disconnections. However, there may still be an EtCO2 reading with a partial disconnection. Similarly, the absence of EtCO2 due to circulatory arrest or a low reading due to a small leak in the sampling line can be mistaken for a disconnection of the breathing system.
Further reading
1. Recommendations for standards of monitoring during anaesthesia and recovery, 4th ed. Association of Anaesthetists of Great Britain and Ireland, 2007. http://www.aagbi.org/publications/guidelines/docs/
standardsofmonitoring07.pdf.
2. Brockwell RC, Andrews JJ. Inhaled anesthetic delivery systems. In: Miller’s anesthesia, Volume 1, 7th ed. Miller RD, Ed. Philadelphia, USA: Churchill Livingstone, 2010; Chapter 25: 667-710.